Display: Vacuum Fluorescent Indicators: Alarm, Alert, Charge, Silenced Flow Rate: 150 ml/min, ± 20 CO2: Measurement: Non-Dispersive IR absorption Measurement Range: 0-10% C02 Units: User-selectable: mmHg, Vol. % kPa Display Range: 0-100 mmHg or 0-10.0% C02 or 0-13.3 kPa Accuracy: ±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10% CO2) Resolution: 1 mmHg; 0.1%; 0.1 kPa Calibration: 2 point Cal N20 Compensation: User-selectable 40% (default=OFF) Response Time: 375 msec to 90% of value Averaging: 4-breath average, or breath-to-breath selectable Respiration Rate: Range: 0-150 bpm Averaging: 4 breath average Accuracy: ± 1 bpm SpO2: Range: 0-100% Functional SpO2 Accuracy: ±2% at 70-100%: <70% unspecified Averaging: 2, 4, 8, or 16-beat averaging (default: 2-beat) Pulse Rate: Range: 30-254 bpm Accuracy: Greater of ±2 bpm or ±2% Averaging: 8 Pulse Tone: Pitch corresponds Sp02 value Volume adjustable or OFF Power Requirements: AC Charger Input 115V-, 60 Hz Battery: Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous- use life 2 hours approximate, maximum full charge time 4 hours Analog Output: Any four of: C02 waveform, Plethysmogram, End Tidal C02, Resiration Rate, Inspired C02, Sp02, Pulse Rate, 0V or 1V Cal Signal Range: 0-1 VDC full scale Serial Output: RS232C: tabular trends, patient data For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec. Dimensions: 3.5"Hx10.0"Wx5.5"D (88.9 mm H x 254mm W x 139.7mm D)